The pervasive contamination of dialysate with microplastics (MPs, <5 mm) and nanoplastics (NPs, <1 μm) presents a critical iatrogenic risk in hemodialysis (HD). This life-sustaining treatment represents a significant exposure route due to its dependence on high-volume dialysate preparation (300–600 L/patient/week) and extensive polymer-based circuitry [1]. The patient’s anephric state enables bioaccumulation with negligible renal clearance of these particulates (Fig. 1).
Primary contamination originates from the reverse osmosis (RO) water used for dialysate. While effective for larger particles, RO removal diminishes for NPs (<255 nm). Polyamide membranes themselves shed NPs via abrasion and chemical degradation [2]. Chlorine breakthrough can oxidize the polymer matrix, accelerating colloidal nanoplastic release into product water, with endogenous contributions quantified at approximately 2.08 μg/L [1]. Feed water is also a source, as treated drinking water contains 6.40 to 8.80 MP particles/L [3]. A secondary critical source is dialysate concentrates. Recent analyses confirm they are a direct vector, with detected MPs including polyethylene, polyvinyl chloride (PVC), and ethylene-vinyl acetate polymers, consistent with packaging materials [4].
A secondary, endogenous source is the extracorporeal blood and dialysate circuits, comprising di(2-ethylhexyl) phthalate (DEHP)-plasticized PVC tubing, polycarbonate connectors, and polysulfone dialyzer membranes. These materials are degraded by roller pump heads and pressure fluctuations, releasing particles directly into the blood path [1]. Concurrently, plasticizers such as DEHP leach into the blood circuit, with intradialytic concentrations reaching 0.2 mg/kg of body mass [5]. Notably, while MP concentrations in HD and peritoneal dialysis fluids are similar, the particles in HD solutions are significantly larger [4].
The toxicological implications for HD patients are direct and multifaceted. The nanoscale size of NPs facilitates translocation across biological barriers. This risk is supported by evidence of significant MPs and NPs bioaccumulation in human organs, with the brain exhibiting particularly high concentrations [6]. At a cellular level, NPs can induce mitochondrial dysfunction, generating reactive oxygen species that activate the NOD-like receptor family pyrin domain containing 3 inflammasome and directly worsen chronic inflammation in HD patients [7]. This cellular damage is driven by well-characterized pathways of oxidative stress and programmed cell death. Furthermore, NPs in dialysate may act as carriers for adsorbed contaminants [8]. The leached DEHP metabolites function as endocrine-disrupting chemicals. Critically, the temporal correlation between peak serum plasticizer concentrations (approximately 6 hours post-dialysis) and the window of highest risk for sudden cardiac death suggests a plausible pathophysiological link [9].
Critical knowledge gaps hinder full risk assessment. Standardized methods are needed to quantify NPs in dialysate and blood. Furthermore, studies are urgently required on MP removal in water treatment and whether online substitution fluids in hemodiafiltration act as direct intravenous particle infusions [1,10].
While in vivo toxicological evidence is strong, a critical gap between research and clinical practice remains. Prospective longitudinal clinical studies are required to directly link internal MP/NP body burden in HD patients with hard endpoints like cardiovascular mortality and the accelerated loss of residual renal function [8,9], a concern directly supported by in vivo evidence that MP/NP exposure induces glomerular damage, tubular injury, and interstitial fibrosis [8]. From a clinical perspective, research must prioritize the development of alternative, non-reactive materials for dialyzers and circuitry. This is supported by data showing microfiltration membranes achieve higher MP removal than adsorption-only systems [3]. Finally, regulatory frameworks for medical devices must account for chronic exposure to particulates from extracorporeal circuits.
The environmental impact of HD is substantial, creating a cycle where medical practice contributes to the pollution that feeds it. A standard HD center acts as a point-source emitter, discharging an estimated 8.4 × 1011 MP particles annually into wastewater via spent dialysate [1], a mass equivalent to roughly 50 kg of plastic waste. This effluent contains persistent polymer fragments and additive residues like DEHP. Furthermore, the RO water treatment systems themselves concentrate removed MPs into sludge (8,425–11,033 particles/kg), creating a secondary pollution hotspot [3]. This cycle highlights a critical externality, as the sector’s emissions contaminate the broader water matrix that supplies the dialysis treatment, thereby inadvertently increasing the toxicological burden on its own vulnerable patient population.
In conclusion, HD is a significant convergence point for human and environmental MP/NP exposure. Mitigation requires a focused approach: enhanced dialysate quality monitoring and DEHP-free circuits in the short term; regulatory action on particulate limits in the mid-term; and the development of closed-loop systems and next-generation biocompatible materials in the long term. Addressing this issue is imperative to reducing an unquantified iatrogenic risk to a vulnerable patient population.

Notes

Conflicts of interest

All authors have no conflicts of interest to declare.

Data sharing statement

No datasets were generated or analyzed during the current study.

Authors’ contributions

Conceptualization: MB, FT

Supervision: MB

Formal analysis, Resources: FT

Writing–original draft: MB, FT

Writing–review & editing: FT

All authors read and approved the final manuscript.

Figure 1.

The lifecycle of microplastic (MP) and nanoplastic (NP) contamination in hemodialysis.

DEHP, di(2-ethylhexyl) phthalate.
j-krcp-25-473f1.jpg

References

1. Passos RS, Davenport A, Busquets R, et al. Microplastics and nanoplastics in haemodialysis waters: emerging threats to be in our radar. Environ Toxicol Pharmacol 2023;102:104253.
crossref
2. Soice NP, Greenberg AR, Krantz WB, Norman AD. Studies of oxidative degradation in polyamide RO membrane barrier layers using pendant drop mechanical analysis. J Membr Sci 2004;243:345–355.
crossref
3. Li Y, Meng Y, Qin L, et al. Occurrence and removal efficiency of microplastics in four drinking water treatment plants in Zhengzhou, China. Water (Basel) 2024;16:131.
crossref
4. Kara E, Konur K, Koca YŞ, Aytan Ü. Unveiling hidden contaminants: a systematic quantification and characterization of microplastics in hemodialysis and peritoneal dialysis fluids. BMC Nephrol 2025;26:359.
crossref pmid pmc pdf
5. Fernandes BL, Maria F, Nakao LS, Ingberman M, Cubas MR. Determination of di(2-ethylhexyl) phthalate released in the blood from polyvinyl chloride line during the hemodialysis procedure. Matéria (Rio J) 2018;23:1–7.
crossref
6. Nihart AJ, Garcia MA, El Hayek E, et al. Bioaccumulation of microplastics in decedent human brains. Nat Med 2025;31:1114–1119.
crossref pmid pmc pdf
7. Alijagic A, Hedbrant A, Persson A, et al. NLRP3 inflammasome as a sensor of micro- and nanoplastics immunotoxicity. Front Immunol 2023;14:1178434.
crossref pmid pmc
8. Verzola D, Rumeo N, Alberti S, et al. Coexposure to microplastic and bisphenol A exhacerbates damage to human kidney proximal tubular cells. Heliyon 2024;10:e39426.
crossref pmid pmc
9. Tereshchenko LG, Posnack NG. Does plastic chemical exposure contribute to sudden death of patients on dialysis? Heart Rhythm 2019;16:312–317.
crossref pmid pmc
10. Chaoui L, Chouati T, Zalegh I, et al. Identification and assessment of antimicrobial resistance bacteria in a hemodialysis water treatment system. J Water Health 2022;20:441–449.
crossref pmid pdf


ABOUT
BROWSE ARTICLES
EDITORIAL POLICY
FOR CONTRIBUTORS
Editorial Office
#301, (Miseung Bldg.) 23, Apgujenog-ro 30-gil, Gangnam-gu, Seoul 06022, Korea
Tel: +82-2-3486-8736    Fax: +82-2-3486-8737    E-mail: registry@ksn.or.kr                

Copyright © 2026 by The Korean Society of Nephrology.

Developed in M2PI

Close layer