Hypersensitivity reactions to synthetic membranes continue to be observed [
The aim of our study was to evaluate small- and mid-sized molecule clearance and albumin leakage in postdilution OL-HDF using the ACT dialyzer.
Nineteen adult patients with stage-5D kidney disease were enrolled in this study. The inclusion criteria were as follows: older than 18 years, completion of three 4-hour sessions weekly for at least 6 months, blood flow > 400 mL/min in the regular sessions, absence of hospital admission in the 4 weeks prior to the study and signed informed consent. Patients who failed to meet the inclusion criteria were excluded. Details of the dialyzer (Solacea dialyzer®, Nipro Medical Corp., Doral, FL, USA) are as follows: membrane surface area, 2.1 m2; KUF, 76 mL/hr/mmHg; inner diameter of hollow fiber, 200 µm; and membrane thickness, 25 µm. Ultrafiltration was prescribed according to the patient’s needs, and 24 L of substitution volume was programmed. The prescribed dialysis features included a 4-hour duration, blood flow of 400 mL/min, and dialysate flow of 700 mL/min.
Pre- and postdialysis blood samples were collected at the mid-week dialysis session. Removal of urea (60 Da), creatinine (113 Da), β2-microglobulin (11.8 kDa), cystatin C (13 kDa), myoglobin (17.2 kDa), and prolactin (23 kDa) was estimated using the reduction ratio (RR) as follows.
(1) RR = (Cpre – Cpost)/Cpre, where Cpre and Cpost are the pre- and posttreatment concentrations, respectively. Posttreatment concentrations of mid-size molecules were corrected for the haemoconcentration using the Bergström and Wehle formula [
β2 microglobulin was measured using a nephelometric immunoassay, and myoglobin and prolactin were assessed via electrochemiluminescence. Albumin was measured using an autoanalyzer. Estimated albumin leakage (EAL) was estimated according to the following formula.
(2) EAL = [15 × (C0 + C15)/2 + 15 × (C15 + C30)/2 + 30 × (C30 + C60)/2 + C60 × (C60 + C120)/2 + C120 × C120]/240 × UF + Sust + (Qd × 240/1,000)], where C is the albumin concentration in dialysate at the beginning (C0) and also at 15 minutes (C15), 30 minutes (C30), 60 minutes (C60), and 120 minutes (C120) (mg/L); UF, ultrafiltration (L); Sust, substitution volume (L); and Qd, dialysate flow (mL/min).
Written informed consent was obtained from all patients. All procedures were in accordance with the Declaration of Helsinki and its revisions. Descriptive results are expressed as the mean ± standard deviation for normally distributed continuous variables and the median and interquartile ranges for non-normally distributed continuous variables. Categorical variables are reported as percentages. All analyses were performed using IBM SPSS for Mac version 20.0 (IBM Corp., Armonk, NY, USA).
The baseline characteristics of the 19 patients included are shown in
Previous
Our study had several limitations. First, this study was developed as an acute study that focused on the results of one dialysis session. Therefore, prospective studies are needed to assess the clinical impact of the use of ACT membranes. Despite the small sample size and absence of a control group, our results are in line with other authors’ [
In conclusion, ACT shows excellent behavior in OL-HDF. Although more prospective studies are needed, according to its clearance results and amount of albumin leakage, this dialyzer is not only an alternative for patients with an allergy to synthetic membranes but is also a good option for nonallergic hemodialysis patients.
All authors have no conflicts of interest to declare.
Conceptualization: NM, AV
Investigation: LC, AV
Methodology: AS
Writing–original draft: AS
Writing–review & editing: LC
All authors read and approved the final manuscript.
Baseline characteristics of the included patients
Characteristic | Data |
---|---|
Age (yr) | 55.0 ± 17.3 |
Male sex | 13 (68.4) |
Dry weight (kg) | 65.4 ± 14.2 |
Serum albumin (g/dL) | 4.1 ± 0.4 |
Hemoglobin (g/dL) | 11.2 ± 1.2 |
Data are expressed as mean ± standard deviation or number (%).
The performance of asymmetric cellulose triacetate dialyzer
Performance | Data |
---|---|
Total convective volume |
27.4 ± 3.4 |
Dialysance (Kt/V) | 1.9 ± 0.4 |
Reduction ratio (%) | |
Urea | 83.7 ± 5.2 |
Creatinine | 76.4 ± 5.3 |
β2 microglobulin | 79.3 ± 4.7 |
Cystatin C | 77.3 ± 4.7 |
Myoglobin | 76.6 ± 5.4 |
Prolactin | 73.7 (67.3–77.5) |
Estimated albumin loss (mg/session) | 481.2 (384.8–596.7) |
Data are expressed as mean ± standard deviation or median (interquartile range).
Substitution volume + ultrafiltration.